  • AP Gain (amplifier gain for AP channels, 50x-2500x, default = 500x)
  • LFP Gain (amplifier gain for LFP channels, 50x-2500x, default = 250x)
  • Reference (External, plus channels 37, 76, 113, 152, 189, 228, 265, 304, 341, or 380, default = External)
  • AP Filter Cut (300 Hz, 500 Hz, or 1000 Hz high pass, default = 300 Hz)

Missing feature: For Option 3 and 4 probes, you can select which channels to acquire data from by clicking on those channels and pressing the ENABLE button. For every channel that is enabled (yellow), another channel is will be disabled (red). There will eventually be an option to disable the output of individual channels (using the ON/OFF buttons) so their data doesn’t flow through the signal chain, but this has not been implemented yet.


Probe calibration

You'll have to manually load the probe’s gain calibration settings from the EEPROM. To do this, press the ADC CALIBRATION and GAIN CALIBRATION buttons. This must be done every time the probe is powered on, and can take up to 10 minutes. The GUI will be frozen during this time, so please be patient! You’ll know it’s finished when the button icon changes color.

...